Standard Protocols
Advanced Techniques
Under development
General Advice
Inspiration
Standard Protocols
Advanced Techniques
Under development
General Advice
Inspiration
Contact person at CAMCOR is Kurt Langworthy (office 541 346 3660, ebeam room 541 346 4778)
Basic parameters are:
At high mag (~ 100,000) find a small round object. Turn on wobble (adjust the extent of the wobble) and watch for lateral motion of the small round object as the focal plane is swept up and down. The aperture position is represented by the red dot on the computer screen. Move the position of the aperature one click at a time until lateral motion is stopped.
At high mag (~ 100,000) find a small round object. Change the focus manually. Look for evidence of two focal planes (seperated in height) at right angles to one another (not necessarily aligned with the x and y axis on the screen). If the two focal planes are separated, a round object will look stretched and blurry along one axis while it looks sharp/small along the perpendicular axis.
Repeat steps 1 and 2 several times.
After drawing a pattern in DesignCAD you make a job file in the Nabity software (different computer from the SEM imaging controls). Exposure parameters for the job file are typically