afm
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
afm [2025/10/07 11:50] – [Tuning the AFM to get a good image] ethanminot | afm [2025/10/07 12:02] (current) – [Step-by-step walk through for AC mode imaging] ethanminot | ||
---|---|---|---|
Line 82: | Line 82: | ||
* Move laser toward the tip of the cantilever - the control wheels are labeled LDX and LDY (laser deflection x and y). They are located on the back & right side of the MFP-3D tripod. Use these controls to maximize the ' | * Move laser toward the tip of the cantilever - the control wheels are labeled LDX and LDY (laser deflection x and y). They are located on the back & right side of the MFP-3D tripod. Use these controls to maximize the ' | ||
* Adjust the photodetector (PD) position. The control wheel for the photodetector is located on the left side of the MFP-3D tripod and is labeled PD. Set the PD ' | * Adjust the photodetector (PD) position. The control wheel for the photodetector is located on the left side of the MFP-3D tripod and is labeled PD. Set the PD ' | ||
- | - X Set AC Mode - In main tab of the master panel select 'AC Mode' in the ' | ||
- Tune the AFM | - Tune the AFM | ||
* Open ' | * Open ' | ||
* Set ' | * Set ' | ||
- | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency. | + | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency |
+ | * In the read-out window of the software, you will now see a 1-volt amplitude signal underneath the sum signal. Maximize this amplitude signal by moving the laser along the length of the AFM cantilever. When the amplitude is maximized, the sum signal will be slightly less than its max value. | ||
+ | * Click 'Auto Tune' one more time (or manually adjust the drive amplitude). | ||
- Engage the tip | - Engage the tip | ||
* Set the I gain to 10 | * Set the I gain to 10 | ||
Line 133: | Line 134: | ||
It’s hard to predict a priori whether the best images will be acquired with phase below 90, or above 90. The best imaging regime for a given day depends on tip sharpness, cantilever stiffness, the sample’s mechanical/ | It’s hard to predict a priori whether the best images will be acquired with phase below 90, or above 90. The best imaging regime for a given day depends on tip sharpness, cantilever stiffness, the sample’s mechanical/ | ||
- | ===== Imaging | + | ===== Other imaging |
It is easiest to get a good image on a small scan area (~ 2 micron). A small scan also facilitates verification of tip sharpness. Starting from the default settings you can fine tune the image and then start increasing the scan size. Good settings will minimize ringing and reduce shadows while keeping the scan rate reasonably fast. | It is easiest to get a good image on a small scan area (~ 2 micron). A small scan also facilitates verification of tip sharpness. Starting from the default settings you can fine tune the image and then start increasing the scan size. Good settings will minimize ringing and reduce shadows while keeping the scan rate reasonably fast. | ||
Line 143: | Line 144: | ||
*Set-point amplitude 0.75 V | *Set-point amplitude 0.75 V | ||
- | **Adjustments to these settings** | ||
- | |||
- | |||
**Using nanotubes as a diagnostic tool** | **Using nanotubes as a diagnostic tool** |
afm.1759863055.txt.gz · Last modified: 2025/10/07 11:50 by ethanminot