afm
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
afm [2025/09/04 11:58] – ethanminot | afm [2025/10/07 12:02] (current) – [Step-by-step walk through for AC mode imaging] ethanminot | ||
---|---|---|---|
Line 63: | Line 63: | ||
===== Step-by-step walk through for AC mode imaging ===== | ===== Step-by-step walk through for AC mode imaging ===== | ||
- | //If a tip change is needed - see the [[AFM tip change information]]// | + | //If a tip change is needed - see the // |
- | - Sign into the log book (black ring binder) on the table next to the AFM. | + | - Sign into the log book (black ring binder) on the table next to the AFM. If Jesse from Inpria used the AFM last, you'll need to mount the standard AFM tip back in the machine: [[AFM tip change information]] |
- Start your RELMS reservation on the [[https:// | - Start your RELMS reservation on the [[https:// | ||
- Turn on the laser - Key switch on the AFM computer | - Turn on the laser - Key switch on the AFM computer | ||
Line 82: | Line 82: | ||
* Move laser toward the tip of the cantilever - the control wheels are labeled LDX and LDY (laser deflection x and y). They are located on the back & right side of the MFP-3D tripod. Use these controls to maximize the ' | * Move laser toward the tip of the cantilever - the control wheels are labeled LDX and LDY (laser deflection x and y). They are located on the back & right side of the MFP-3D tripod. Use these controls to maximize the ' | ||
* Adjust the photodetector (PD) position. The control wheel for the photodetector is located on the left side of the MFP-3D tripod and is labeled PD. Set the PD ' | * Adjust the photodetector (PD) position. The control wheel for the photodetector is located on the left side of the MFP-3D tripod and is labeled PD. Set the PD ' | ||
- | - X Set AC Mode - In main tab of the master panel select 'AC Mode' in the ' | ||
- Tune the AFM | - Tune the AFM | ||
* Open ' | * Open ' | ||
* Set ' | * Set ' | ||
- | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency. | + | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency |
+ | * In the read-out window of the software, you will now see a 1-volt amplitude signal underneath the sum signal. Maximize this amplitude signal by moving the laser along the length of the AFM cantilever. When the amplitude is maximized, the sum signal will be slightly less than its max value. | ||
+ | * Click 'Auto Tune' one more time (or manually adjust the drive amplitude). | ||
- Engage the tip | - Engage the tip | ||
* Set the I gain to 10 | * Set the I gain to 10 | ||
Line 95: | Line 96: | ||
* Lower the 'Set Point Voltage' | * Lower the 'Set Point Voltage' | ||
* When the Z-voltage is above its midpoint, use the thumb wheel to lower the tripod leg until the Z-voltage is below the midpoint. | * When the Z-voltage is above its midpoint, use the thumb wheel to lower the tripod leg until the Z-voltage is below the midpoint. | ||
- | * Lower the 'Set Point Voltage' | + | * Lower the 'Set Point Voltage' |
* After you achieve a hard stop, disengage the tip by clicking ' | * After you achieve a hard stop, disengage the tip by clicking ' | ||
- Close AFM Hood gentle (don't bump the machine). | - Close AFM Hood gentle (don't bump the machine). | ||
- Check the Amplitude and Phase of the cantilever vibration. Re-tune if needed. | - Check the Amplitude and Phase of the cantilever vibration. Re-tune if needed. | ||
- | - Set image details | + | - Set imaging parameters |
- | - Scan the sample. Clicking 'frame up' or 'frame down' will start a scan | + | - Do your first test scan the sample. Clicking 'frame up' or 'frame down' will start a scan. You should do a test scan before you spend much time searching for a specific location on your sample. |
- | - Observe the quality of the scan. Look for phase jumps. Look for parachuting. Look for the tip loosing contact with the surface. If needed, you can change | + | - Observe the quality of the scan. See the section below called " |
+ | - If you got a good image, you are ready for the rest of your session. If you need the microscope to search a large area, raise the tip by one " | ||
Line 118: | Line 120: | ||
- Leave controller and PC running unless expecting a power outage | - Leave controller and PC running unless expecting a power outage | ||
- | ===== Imaging rules of thumb ===== | + | ===== Tuning the AFM to get a good image ===== |
- | It is easiest to get a good image on a small scan area (~ 2 micron). A small scan also facilitates verification of tip sharpness. Starting from the default settings you can fine tune the image and then start increasing the scan size. Good settings will minimize ringing and reduce shadows while keeping the scan rate reasonably fast. | + | |
- | **Beginner settings** | + | Check the **scan rate**. Is it less than 15 microns/ |
- | *Scan size 4 micron | + | |
- | *Scan rate < 15 micron/s | + | |
- | *Integral gain 10 | + | |
- | *Free amplitude 1 V (corresponds to a cantilever motion of ~ 100 nm) | + | |
- | *Set-point amplitude 0.75 V | + | |
- | + | ||
- | **Adjustments to these settings** | + | |
When you withdraw from the surface, check the free-air amplitude is the same as you when you first tuned the tip. If **free-air amplitude has drifted**, you can manually change the drive amplitude. | When you withdraw from the surface, check the free-air amplitude is the same as you when you first tuned the tip. If **free-air amplitude has drifted**, you can manually change the drive amplitude. | ||
Line 134: | Line 128: | ||
Sometimes the image is improved by withdrawing and re-running the autotune procedure (**resonant frequency might have changed**). The autotune procedure will make changes to both the drive amplitude and drive frequency. | Sometimes the image is improved by withdrawing and re-running the autotune procedure (**resonant frequency might have changed**). The autotune procedure will make changes to both the drive amplitude and drive frequency. | ||
- | Sometimes the image is improved by lowering the set-point amplitude a few clicks. For example, this might fix **parachuting**. Minor changes to set-point amplitude can be made in real time, during imaging. | + | Sometimes the image is improved by lowering the set-point amplitude a few clicks. For example, this might fix **parachuting**. Minor changes to set-point amplitude can be made in real time, during imaging. The set-point should be low enough that the tip stays in contact with the surface. However, don't make the set-point too low. The ideal set-point amplitude is typically about two clicks lower than loosing contact with the surface. |
- | Another thing to try is a slower | + | If you notice |
- | If you notice **phase jumping** (a jump from below 90 degree to above 90 degrees), you should try adjusting | + | It’s hard to predict |
+ | |||
+ | ===== Other imaging rules of thumb ===== | ||
+ | It is easiest | ||
+ | |||
+ | **Beginner settings** | ||
+ | *Scan size 4 micron (look at a random small feature on a flat background to verify the sharpness of the tip) | ||
+ | *Scan rate < 15 micron/s | ||
+ | *Integral gain 10 | ||
+ | *Free amplitude 1 V (corresponds to a cantilever | ||
+ | *Set-point amplitude 0.75 V | ||
- | It’s hard to predict a priori whether the best images will be acquired with phase below 90, or above 90. The best imaging regime for a given day depends on tip sharpness, cantilever stiffness, the sample’s mechanical/ | ||
**Using nanotubes as a diagnostic tool** | **Using nanotubes as a diagnostic tool** | ||
Line 183: | Line 186: | ||
- If something is stuck to your AFM tip, it will show up in a Force curve. | - If something is stuck to your AFM tip, it will show up in a Force curve. | ||
- | **To replace a tip** | ||
- | - Invert the AFM head onto the stabilization table. Remove the tip holder by pressing the black rubber button and secure it in the temporary mount by the laser. | ||
- | - Gently unscrew the clamp holding the tip down with a Phillips screwdriver. You don't want to completely unscrew it, just loosen it enough to remove the tip. | ||
- | - Grab the old tip by its sides with a pair of tweezers and pull it away from the clamp. It should slide out freely. | ||
- | - Remove a new cantilever from the box with a pair of plastic tweezers using a twisting motion - be careful not to touch the end with the tip! Grab the new cantilever with your tweezers so the end you want to insert into the tip holder is pointing away from you and the tip is pointing upwards. | ||
- | - Slide the tip between the tip holder and the clamp (probably just as the old one was placed). The cantilever should be placed so it isn't inserted too far or too short. Too far, and the tip will not be well situated in the holder and zeroing the deflection will be impossible (see the figure below). | ||
- | - Once the tip is inserted properly, tighten the clamp and replace the holder in the AFM head. Make sure to rotate the AFM head in the opposite direction you used to invert the head, to avoid twisting the head cable. | ||
- | - Test the tip insertion two ways: i) Make sure the laser spot can be placed on the tip and the deflection can be zeroed without placing the wheels at their maximum range. ii) Auto tune the AFM, and make sure the amplitude and phase plots are free of any irregularities. | ||
- | {{AFMCantelieverPosition.png|}} | ||
===== Trouble shooting ===== | ===== Trouble shooting ===== | ||
AFM doesn' | AFM doesn' |
afm.1757012304.txt.gz · Last modified: 2025/09/04 11:58 by ethanminot