afm
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
afm [2025/07/23 17:59] – ethanminot | afm [2025/10/07 12:02] (current) – [Step-by-step walk through for AC mode imaging] ethanminot | ||
---|---|---|---|
Line 14: | Line 14: | ||
Basic rules about booking time: | Basic rules about booking time: | ||
*People who have scheduled time get priority. | *People who have scheduled time get priority. | ||
- | *Don't block off an entire 9am-5pm workday - leave at least an hour for someone | + | *For OSU internal users, please limit your bookings |
+ | *Only make a booking if you are 90% sure you will use the booking. When you make a booking, everyone else adjusts their schedule to work around your booking. Be respectful of the effort other users are making for you. | ||
Every time you use the AFM, follow these steps: | Every time you use the AFM, follow these steps: | ||
Line 62: | Line 63: | ||
===== Step-by-step walk through for AC mode imaging ===== | ===== Step-by-step walk through for AC mode imaging ===== | ||
- | //If a tip change is needed - see the [[AFM tip change information]]// | + | //If a tip change is needed - see the // |
- | - Sign into the log book (black ring binder) on the table next to the AFM. | + | - Sign into the log book (black ring binder) on the table next to the AFM. If Jesse from Inpria used the AFM last, you'll need to mount the standard AFM tip back in the machine: [[AFM tip change information]] |
- Start your RELMS reservation on the [[https:// | - Start your RELMS reservation on the [[https:// | ||
- Turn on the laser - Key switch on the AFM computer | - Turn on the laser - Key switch on the AFM computer | ||
Line 81: | Line 82: | ||
* Move laser toward the tip of the cantilever - the control wheels are labeled LDX and LDY (laser deflection x and y). They are located on the back & right side of the MFP-3D tripod. Use these controls to maximize the ' | * Move laser toward the tip of the cantilever - the control wheels are labeled LDX and LDY (laser deflection x and y). They are located on the back & right side of the MFP-3D tripod. Use these controls to maximize the ' | ||
* Adjust the photodetector (PD) position. The control wheel for the photodetector is located on the left side of the MFP-3D tripod and is labeled PD. Set the PD ' | * Adjust the photodetector (PD) position. The control wheel for the photodetector is located on the left side of the MFP-3D tripod and is labeled PD. Set the PD ' | ||
- | - X Set AC Mode - In main tab of the master panel select 'AC Mode' in the ' | ||
- Tune the AFM | - Tune the AFM | ||
* Open ' | * Open ' | ||
- | * Set ' | + | * Set ' |
- | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency. | + | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency |
+ | * In the read-out window of the software, you will now see a 1-volt amplitude signal underneath the sum signal. Maximize this amplitude signal by moving the laser along the length of the AFM cantilever. When the amplitude is maximized, the sum signal will be slightly less than its max value. | ||
+ | * Click 'Auto Tune' one more time (or manually adjust the drive amplitude). | ||
- Engage the tip | - Engage the tip | ||
* Set the I gain to 10 | * Set the I gain to 10 | ||
- | * Make the 'Set Point Voltage' | + | * Make the 'Set Point Voltage' |
- | * Click ' | + | * Click ' |
* Lower the tip down towards the sample by shortening the front leg of the MFP-3D tripod. Watch the measured amplitude as you do this. Also watch the camera image. As the surface comes into focus (before the amplitude starts to drop), consider moving to a clean place on your sample (use the x-y stage coarse positioning micrometers). | * Lower the tip down towards the sample by shortening the front leg of the MFP-3D tripod. Watch the measured amplitude as you do this. Also watch the camera image. As the surface comes into focus (before the amplitude starts to drop), consider moving to a clean place on your sample (use the x-y stage coarse positioning micrometers). | ||
- | * If you have a clean landing area for your tip, continue lowering the tip down towards the sample. The amplitude will drop as you near the surface. You will also notice changes in phase. Pay attention to these " | + | * If you have a clean landing area for your tip, continue lowering the tip down towards the sample. The amplitude will drop as you near the surface. You will also notice changes in phase. Pay attention to these "tell-tales". The computer will beep when feedback kicks in to stop the amplitude from dropping below the setpoint. Continue lowering until the Z voltage is just past the middle of its range. (If the Z-voltage is not behaving as expected, click withdraw and check the amplitude and phase). |
* Lower the 'Set Point Voltage' | * Lower the 'Set Point Voltage' | ||
* When the Z-voltage is above its midpoint, use the thumb wheel to lower the tripod leg until the Z-voltage is below the midpoint. | * When the Z-voltage is above its midpoint, use the thumb wheel to lower the tripod leg until the Z-voltage is below the midpoint. | ||
- | * Lower the 'Set Point Voltage' | + | * Lower the 'Set Point Voltage' |
* After you achieve a hard stop, disengage the tip by clicking ' | * After you achieve a hard stop, disengage the tip by clicking ' | ||
- Close AFM Hood gentle (don't bump the machine). | - Close AFM Hood gentle (don't bump the machine). | ||
- Check the Amplitude and Phase of the cantilever vibration. Re-tune if needed. | - Check the Amplitude and Phase of the cantilever vibration. Re-tune if needed. | ||
- | - Set image details | + | - Set imaging parameters |
- | - Scan the sample. Clicking 'frame up' or 'frame down' will start a scan | + | - Do your first test scan the sample. Clicking 'frame up' or 'frame down' will start a scan. You should do a test scan before you spend much time searching for a specific location on your sample. |
+ | - Observe the quality of the scan. See the section below called " | ||
+ | - If you got a good image, you are ready for the rest of your session. If you need the microscope to search a large area, raise the tip by one " | ||
Line 105: | Line 109: | ||
- Click ' | - Click ' | ||
- Open AFM Hood | - Open AFM Hood | ||
- | - Manually retract tip from sample - Give the front thumbwheel | + | - Manually retract tip from sample - turn the front thumbwheel |
- | - Turn off laser - Key on the AFM computer | + | - Turn off laser - Key on the AFM controller. |
- | - Turn off camera light - Switch on the box sitting on top of the AFM | + | - Turn off camera light - Switch on the box sitting on top of the AFM enclosure. |
- | - Place MFP-3D onto its shelf holder | + | - Place MFP-3D |
- Remove sample | - Remove sample | ||
- | - Close software | + | |
+ | | ||
+ | - Sign out of the Log Book. | ||
+ | - End the reservation in RELMS. | ||
- Leave controller and PC running unless expecting a power outage | - Leave controller and PC running unless expecting a power outage | ||
- | ===== Imaging | + | ===== Tuning the AFM to get a good image ===== |
+ | |||
+ | Check the **scan rate**. Is it less than 15 microns/ | ||
+ | |||
+ | When you withdraw from the surface, check the free-air amplitude is the same as you when you first tuned the tip. If **free-air amplitude has drifted**, you can manually change the drive amplitude. | ||
+ | |||
+ | Sometimes the image is improved by withdrawing and re-running the autotune procedure (**resonant frequency might have changed**). The autotune procedure will make changes to both the drive amplitude and drive frequency. | ||
+ | |||
+ | Sometimes the image is improved by lowering the set-point amplitude a few clicks. For example, this might fix **parachuting**. Minor changes to set-point amplitude can be made in real time, during imaging. The set-point should be low enough that the tip stays in contact with the surface. However, don't make the set-point too low. The ideal set-point amplitude is typically about two clicks lower than loosing contact with the surface. | ||
+ | |||
+ | If you notice **phase jumping** (a jump from below 90 degree to above 90 degrees), you need to make adjustments to the drive frequency. Phase jumping will wear out the AFM tip (making it blunt). Phase jumping also introduces artifacts in the height data. Make test images with different values of drive frequency such that the free-air phase is 70 degrees, 80 degrees, 100 degree and 110 degrees. To make these test images, you'll need to maintain a constant free-air amplitude by simultaneously adjusting drive amplitude. By doing this, you are searching for imaging parameters for which the cantilever oscillations are most stable. | ||
+ | |||
+ | It’s hard to predict a priori whether the best images will be acquired with phase below 90, or above 90. The best imaging regime for a given day depends on tip sharpness, cantilever stiffness, the sample’s mechanical/ | ||
+ | |||
+ | ===== Other imaging | ||
It is easiest to get a good image on a small scan area (~ 2 micron). A small scan also facilitates verification of tip sharpness. Starting from the default settings you can fine tune the image and then start increasing the scan size. Good settings will minimize ringing and reduce shadows while keeping the scan rate reasonably fast. | It is easiest to get a good image on a small scan area (~ 2 micron). A small scan also facilitates verification of tip sharpness. Starting from the default settings you can fine tune the image and then start increasing the scan size. Good settings will minimize ringing and reduce shadows while keeping the scan rate reasonably fast. | ||
**Beginner settings** | **Beginner settings** | ||
- | *Scan size 2 micron (look at a random small feature on a flat background to verify the sharpness of the tip) | + | *Scan size 4 micron (look at a random small feature on a flat background to verify the sharpness of the tip) |
*Scan rate < 15 micron/s | *Scan rate < 15 micron/s | ||
*Integral gain 10 | *Integral gain 10 | ||
Line 123: | Line 144: | ||
*Set-point amplitude 0.75 V | *Set-point amplitude 0.75 V | ||
- | Sometimes the image is improved by lowering the set-point amplitude a few clicks. For example, this might fix parachuting. | ||
- | |||
- | Sometimes the image is improved by withdrawing and re-running the autotune procedure (the resonant frequency might have changed). | ||
- | |||
- | When you withdraw from the surface, check the free air amplitude is the same as you when you first tuned the tip. | ||
- | |||
- | Another thing to try is a slower scan rate. The price you pay is scan time. However, I've found that you actually save time (and headache) by taking a single high quality slow scan rather than a bunch of quick ones with little parameter adjustments in between. I find that adjusting the rate so that the scan speed is <10 micron/sec works well in nearly all cases. | ||
**Using nanotubes as a diagnostic tool** | **Using nanotubes as a diagnostic tool** | ||
Line 172: | Line 186: | ||
- If something is stuck to your AFM tip, it will show up in a Force curve. | - If something is stuck to your AFM tip, it will show up in a Force curve. | ||
- | **To replace a tip** | ||
- | - Invert the AFM head onto the stabilization table. Remove the tip holder by pressing the black rubber button and secure it in the temporary mount by the laser. | ||
- | - Gently unscrew the clamp holding the tip down with a Phillips screwdriver. You don't want to completely unscrew it, just loosen it enough to remove the tip. | ||
- | - Grab the old tip by its sides with a pair of tweezers and pull it away from the clamp. It should slide out freely. | ||
- | - Remove a new cantilever from the box with a pair of plastic tweezers using a twisting motion - be careful not to touch the end with the tip! Grab the new cantilever with your tweezers so the end you want to insert into the tip holder is pointing away from you and the tip is pointing upwards. | ||
- | - Slide the tip between the tip holder and the clamp (probably just as the old one was placed). The cantilever should be placed so it isn't inserted too far or too short. Too far, and the tip will not be well situated in the holder and zeroing the deflection will be impossible (see the figure below). | ||
- | - Once the tip is inserted properly, tighten the clamp and replace the holder in the AFM head. Make sure to rotate the AFM head in the opposite direction you used to invert the head, to avoid twisting the head cable. | ||
- | - Test the tip insertion two ways: i) Make sure the laser spot can be placed on the tip and the deflection can be zeroed without placing the wheels at their maximum range. ii) Auto tune the AFM, and make sure the amplitude and phase plots are free of any irregularities. | ||
- | {{AFMCantelieverPosition.png|}} | ||
===== Trouble shooting ===== | ===== Trouble shooting ===== | ||
AFM doesn' | AFM doesn' |
afm.1753318798.txt.gz · Last modified: 2025/07/23 17:59 by ethanminot