afm
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
afm [2022/07/13 13:21] – dublin | afm [2025/02/09 13:46] (current) – ethanminot | ||
---|---|---|---|
Line 5: | Line 5: | ||
*//See also: [[AFM upkeep]]// | *//See also: [[AFM upkeep]]// | ||
- | New users on the AFM are expected to get AFM training and then pass the {{:: | + | New users of the AFM are expected to get AFM training and then pass the {{:: |
To find answers to some of the quiz questions, the "AFM Manualette" | To find answers to some of the quiz questions, the "AFM Manualette" | ||
*T: | *T: | ||
Line 65: | Line 65: | ||
- Place the sample to be imaged on the tray. Make sure that the stage x-y control thumbscrews are centered. | - Place the sample to be imaged on the tray. Make sure that the stage x-y control thumbscrews are centered. | ||
- Make sure the vibration isolation stage is on and isolation is enabled. | - Make sure the vibration isolation stage is on and isolation is enabled. | ||
- | - Raise the legs on the MFP-3D tripod to ensure the tip does not smash into the sample. | + | - Raise the legs on the MFP-3D tripod |
- Set the MFP-3D over sample | - Set the MFP-3D over sample | ||
- Align Laser: | - Align Laser: | ||
- | * Turn on the camera - Click the lower left icon with a picture of a TV on it | + | * Turn on the camera - Click the lower left icon with a picture of a camera |
* Turn on the camera light - Switch on the box which sits on top of the AFM | * Turn on the camera light - Switch on the box which sits on top of the AFM | ||
* Align camera on cantilever - Two knobs sticking up at the very rear of the MFP-3D | * Align camera on cantilever - Two knobs sticking up at the very rear of the MFP-3D | ||
Line 78: | Line 78: | ||
- X Set AC Mode - In main tab of the master panel select 'AC Mode' in the ' | - X Set AC Mode - In main tab of the master panel select 'AC Mode' in the ' | ||
- Tune the AFM | - Tune the AFM | ||
- | * X Open ' | + | * Open ' |
- | * X Set ' | + | * Set ' |
* Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency. | * Click the 'Auto Tune' button and wait for tuning to finish. Software will set drive frequency. | ||
- Engage the tip | - Engage the tip | ||
- | * X Set the I gain to 10 | + | * Set the I gain to 10 |
- | * Make the 'Set Point Voltage' | + | * Make the 'Set Point Voltage' |
* Click ' | * Click ' | ||
- | * Lower the tip (tripod thumbwheel) while watching the amplitude. The amplitude will drop as you near the surface. The computer will beep when feedback kicks in to stop the amplitude from dropping below the setpoint. Continue lowering until the Z voltage is in the middle of its range. | + | * Lower the tip (tripod thumbwheel |
+ | * If the Z-voltage goes down instead of up, re-tune. If that doesn' | ||
* Lower the 'Set Point Voltage' | * Lower the 'Set Point Voltage' | ||
* Lower the tip until the 'Z voltage' | * Lower the tip until the 'Z voltage' | ||
Line 100: | Line 101: | ||
- Manually retract tip from sample - Give the front thumbwheel a few clockwise twists | - Manually retract tip from sample - Give the front thumbwheel a few clockwise twists | ||
- Turn off laser - Key on the AFM computer | - Turn off laser - Key on the AFM computer | ||
- | - Turn off camera - Close camera window | ||
- Turn off camera light - Switch on the box sitting on top of the AFM | - Turn off camera light - Switch on the box sitting on top of the AFM | ||
- Place MFP-3D onto its shelf holder | - Place MFP-3D onto its shelf holder | ||
- Remove sample | - Remove sample | ||
+ | - Close software and log out | ||
- Leave controller and PC running unless expecting a power outage | - Leave controller and PC running unless expecting a power outage | ||
===== Imaging rules of thumb ===== | ===== Imaging rules of thumb ===== | ||
Line 110: | Line 111: | ||
**Beginner settings** | **Beginner settings** | ||
*Scan size 1 micron | *Scan size 1 micron | ||
- | *Scan rate 1 Hz | + | *Scan rate < 15 micron / s |
*Integral gain 10 | *Integral gain 10 | ||
*Free amplitude 1 V (~ 100 nm) | *Free amplitude 1 V (~ 100 nm) |
afm.1657743697.txt.gz · Last modified: 2022/07/13 13:21 by dublin