Starting the Diffusion Pump
-
Turn on the power to the surge protectors
Make sure the mechanical pump is on
Make sure all valves are closed
Open the foreline valve to back the diff pump
flip on the power switch to the diff pump on the front panel
Loading a Sample
Make sure the high vacuum valve and the roughing valve are closed
Open the vent valve
Open the tank of nitrogen gas attached to the vent. The nitrogen is used to speed up the venting process.
Wait until you see that the bell jar seal has been broken or the pressure in the chamber has reached 760 torr.
Close vent so nitrogen doesn't leak into room
Lift the bell jar by using the switch on the front panel
Put sheet metal plate back on
Power on & test that the shutter mechanism is working. Then power it off.
Lower the bell jar and take care that the jar is seated properly
Pumping down
Close the vent valve
Close the foreline valve
Open the roughing valve.
Important: mechanical pump should never pump with the foreline & roughing valves both open
While you are roughing the chamber the pressure in the foreline may rise. If the pressure in the foreline (needle gauge) exceeds 100mTorr close the roughing valve and open the foreline valve. Once the pressure in the foreline is back to normal, close the foreline valve again and open the roughing valve to resume pumping.
Wait until the pressure in the chamber is around 5×10-2 Torr
Close the roughing valve
Open the foreline valve
Turn off the ion gauge, if on
Pour liquid nitrogen into funnel behind the bell jar
Slowly open the High Vacuum valve making sure not to let the foreline pressure go above 100 mTorr
After the high vacuum valve has been completely opened, wait 5 minutes then turn on the ion gauge
go to pressure display channel 1
press ems to turn off/on
Note: Don't let the ion gauge be turned on when the pressure is greater than 10-4 Torr
If you think there's a problem with the ion gauge press 'degas' no pressure display channel 1. This will try to evaporate contaminants out of the gauge. Turn off 'degas' before continuing.
Wait until the pressure in the chamber has reached about 10-6 Torr
Metal deposition
Turn on the water supply to the boat electrodes, first the return then the supply
Program in the correct density and Z-factor into the film thickness gauge. You may have to do this more than once if you have more than one metal.
Turn on the power supply to the boat
Turn up the voltage on the power supply in small increments, and wait about a minute in between increments. Keep increasing the voltage until the desired deposition rate is achieved.
Turn on power to shutter mechanism
Acting quickly, open the shutter and press 'zero' to zero the thickness gauge.
Close the shutter when you've reached the right thickness.
Turn off shutter power
When finished slowly turn down power to boat
Turn off power to boat
Removing your Sample
Removing a sample is just like loading a sample except when you are finished leave the chamber in a rough vacuum state.
Shutting Down the Diffusion Pump
If no one will need to use the thermal evaporator for a while it maybe a good idea to shut down the machine.
Turn off the power to the diff pump by flipping the switch on the front panel
Leave the water on and the mechanical pump on until the diffusion pump has cooled down.
Once it has cooled close the foreline valve.
Turn off the water by closing the supply then the return
Turn off the power on the surge protectors, this will turn everything off.